Operating instructions HE lite
© 2017 – Pfeuffer GmbH Revision 9 Page 7 of 28
5 Operation/Use
5.1 General/Operational control/Display/Notice:
The HE lite switches off automatically after 30 seconds, if
no button was pressed in the meantime. The display has
four lines. In the last line the current function of the
buttons is indicated in white letters on black background.
In the picture on the right, arrow to the left, measure and
off.
The instrument is equipped with three operational
buttons. The arrow keys and the On/Off-button have
additionally the function indicated on the display. The
appropriate function is indicated above each button.
6 Moisture measurement
6.1 Filling the measuring cell and grinding
The measuring cell is equipped with
special ribbing allowing the crushing of
grain and oil seeds.
Attention!
Foreign elements, such as small
stones, straw, husks, weed seeds,
green kernels or in case of corn, stalks
should be removed from the sample
prior to the measurement.
Open the top unit of the measuring cell
anticlockwise.
Fill the filling measure (11 ml) inside the
top unit of the measuring cell to the brim
with the cleaned sample.
Empty the contents of the filling measure
evenly
into the bottom unit of the
measuring cell.