- 31 -
(Note)
1. Customer’s program will be deleted all at once if you perform “Delete All User Data” (Function
#925) on Select Tools Svc screen, but calibration value will not be initialized.
2. How to perform “Delete All User Data” (Function #925)
(1) In Top screen, select “Tools” on MENU bar to display Select Tools screen.
(2) Select “Svc” on MENU bar to display Password screen.
(3) Input Service code “384” and select “OK” on MENU bar.
(4) Select “Calibration 2“ and select “ OK” on MENU bar.
(5) Input Function number “925” besides “Function” column.
(6) Select “Tools” on MENU bar to display Select Tools screen.
(7) Select “Svc” on MENU bar and perform “Delete All User Data”.
(8) Select “OK” on MENU bar to initialize data.
(9) Turn main power off and on.
3. Initial setting values after Function #925 is performed
Stand-by temperature : 25
Illuminance : 0 (Off)
Humidity : 60%
Temperature zero calibration : 500
Humidity zero calibration : 510
Temperature span calibration : 505
Humidity span calibration : 560
Automatic defrosting : 2
(Evaporator and evaporator for de-humidity; 3:00a.m. start
Evaporator for de-humidity; 3:00p.m. start)
Interval of log correction : 6 minutes
Electric valve : 500 (Automatic)
LED back color : 1 (Blue)
Compressor running mode : 0 (Automatic)
Type of model : 1 (MLR-352H)
Communication speed (DAQ) : 0 (2400bps)
Communication ID (DAQ) : 0 (OFF)
Communication mode : 0 (Local)
Buzzer for completing program run: 2 (None)
Temperature alarm : Set temperature +/-2.5
Humidity alarm : Set humidity +/-10%
Alarm delay time : 15 minutes
Ring back time : 30 minutes
Delay time for door alarm : 2 minutes
Ring back time for door alarm : 0 (OFF)
Key lock : Release
Key lock password : 0000
Offset for solenoid valve : 50