- 41 -
3
Normal maintenance
(a)Clean-up strainer
1) Arrange equipments.
2) Turn off main power switch and close water feed
 service valve.
3) Disconnect cupper tube and make sure to secure
 strainer with 2pcs. of adjustable wrench.
4) Remove cover from strainer and make sure to
secure valve joint with 2pcs. of adjustable
wrench.

5) Loosen and remove strainer with a slotted
screwdriver

6) Clean-up inside of strainer and wash
with water.
7) Follow the procedure 1)~6) on reverse order
to reassemble strainer.

Orifice ȭ4mm
Strainer
Strainer
Cover of strainer