Getting Started
DigiTrak XT Recorder Instructions for Use 1-15
7 Observe the screen for cable and lead errors as follows.
8 When all the leads are properly connected, the ECG waveform appears.
9 Use the arrow buttons to change the gains settings of the display, as shown in the
following table.
NOTE The recording for all resolutions is always made at the 1X gain setting.
10
Proceed to Settings and select Pacer Detector, if applicable.
11 Use the arrow buttons to select the Start tab and press Enter to start the recording.
NOTE If you do not start the recording, the DigiTrak XT automatically starts recording after 30 minutes,
provided there are no lead errors.
During a Recording
During a recording, the DigiTrak XT displays the date, current time, and the time remaining
for the recording as well as any lead or cable error, if present.
Table 1-2 Gain Settings
Gain on screen Millivolt range
0.5X ± 5 peak-to-peak
1X ± 2.5 peak-to-peak
2X ± 1.25 peak-to-peak