Troubleshooting
DigiTrak XT Recorder Instructions for Use 2-3
Noise artifacts on ECG signal Ensure you have prepared the patient’s skin according
to the instructions on page 1-12.
Ensure the electrodes are properly applied to the
patient.
Ensure the leads are making proper contact with the
electrodes.
Ensure patient cable (lead set) is making contact with
the electrodes.
Replace the lead set.
No Lead Connected message Ensure you have prepared the patient’s skin according
to the instructions on page 1-12.
Ensure the electrodes are properly applied to the
patient.
Ensure the leads are making proper contact with the
electrodes.
Ensure patient cable (lead set) is connected.
Replace the lead set.
Leads Error
Ensure you have prepared the patient’s skin according
to the instructions on page 1-12.
Ensure the electrodes are properly applied to the
patient.
Ensure the leads are making proper contact with the
electrodes.
Ensure patient cable (lead set) is connected.
Replace the lead set.
Defective Card message Remove and reinsert battery. Note that when you
restart the recorder, all patient and ECG information is
erased from the memory card and cannot be recovered.
If message still appears, call for service.
Existing ECG in recorder. The
patient’s name is displayed on
the recorder screen. The
recorder does not power on.
Download the ECG data to the Holter application.
Self test failure
Write down the error code.
Restart the recorder.
Call for service.
Table 2-1 Troubleshooting (continued)
Symptom Solution