SERVICE & TECHNICAL INFORMATION 1039055, VER. 06
3.
Connect the pressure gauge from the tool kit to the outlet barb. The flow ball should drop to 0.0
LPM.
4.
If the outlet pressure does not fall between 5.0 and 7.0 psig, perform steps 5 through 14. If the
pressure does fall between 5.0 and 7.0 psig, remove pressure gauge, the test is now complete.
5.
Turn the unit off.
6.
Remove the pressure gauge from the outlet barb.
7.
Remove the front and rear cabinets.
NOTE
Be sure to hold the pressure gauge in a vertical position.