SERVICE & TECHNICAL INFORMATION 1039055, VER. 06
7.3 TROUBLESHOOTING TABLE
SYMPTOM CAUSE VERIFICATION
CORRECTIVE
ACTION
Low pressure
Canister to Compressor
tubing failure
Compressor failure
Valve Solenoid failure
Sieve Canister leaking
Verify there are no
leaks
Verify the tubing
between the canister
and Compressor is
connected
Verify the tubing is
not kinked or
pinched
Verify there are no
leaks in the hose
from Compressor to
Sieve Canister
Verify the
Compressor is
operating correctly
Verify the Valve
Solenoid is working
Verify wiring harness
from the Valve
Solenoid to Main
PCA are connected
Verify there are no
leaks in Sieve
Canister
Repair leaks in
tubing and hoses
Reconnect or
replace the
tubing
Reposition the
tubing
Replace clamps or
replace hose
Replace the
Compressor
Replace the Valve
Body
Reconnect the
wiring harnesses
and retest
If problem
continues, install
old valve body
and new pilot
valves
Replace the Sieve
Canister
Assembly