SERVICE & TECHNICAL INFORMATION 1039055, VER. 06
8.6 O
2
QUICK COUPLER REPLACEMENT
To remove the O
2
Quick Coupler
1.
Remove the Filter Cover. Refer to the Filter Cover Replacement section for more information.
2.
Remove the Inlet Filter. Refer to the Inlet Filter Replacement section for more information.
3.
Remove the Rear Cabinet. Refer to the Rear Cabinet/Power Cord Replacement section for more
information.
4.
Cut the tie wrap securing the tubing to the O
2
Coupler.
5.
Remove the securing nut from the coupler and push the coupler towards the outside of the device
to remove.
To install the O
2
Coupler
1.
Insert the O
2
Coupler through the side wall threads toward the inside of the device ensuring that
the flat “D” portion of the Coupler is aligned with the flat “D” portion of the pass through hole.
2.
Install the securing nut to the O
2
Coupler and tighten.
3.
Install the tubing to the O
2
Coupler.
4.
Secure the tubing to the O
2
Coupler using a tie wrap.
5.
Install the Rear Cabinet. Refer to the Rear Cabinet/Power Cord Replacement section for more
information.
6.
Install the Inlet Filter. Refer to the Inlet Filter Replacement section for more information.
7.
Install the Filter Cover. Refer to the Filter Cover Replacement section for more information.