7.4 SYSTEM PRESSURE TEST TABLE
SYMPTOM CAUSE VERIFICATION
CORRECTIVE
ACTION
System Pressure Test
Abnormal reading
High-pressure
reading above 25
psig @ 60 Hz @ 5
LPM flow.
Check oxygen
percentage at 5
LPM. If Low?
Check for proper
valve cycling
Replace Sieve
Canister
Fluctuating
pressure
Difference in Peak
pressure is > 1 psig.
Check oxygen
percentage at 5
LPM. If Low?
Check for leaks
Replace Sieve
Canister
Low pressure
reading
Low pressure reading
below 20 psig @ 5
LPM
Check oxygen
percentage at 5
LPM. If Low?
Check for leaks
Check tubing
Check fittings
Check for dirty filters
Check Compressor
for System Pressure
Replace tubing
Replace fittings
Replace filters
Replace
Compressor