Page 33 E
VERFLO
TM
SERVICE & TECHNICAL INFORMATION 1039055, VER. 06
SYMPTOM CAUSE VERIFICATION
CORRECTIVE
ACTION
High pressure
Pressure relief valve
Valve Solenoid failure
Verify that tubing is
not kinked or
pinched
Verify that the tubing
is not damaged
Verify the valve
wiring harness is
connected at the
Main PCA
Verify the
Compressor is
working correctly
Verify the Valve
Solenoid is working
properly
Verify the valve
wiring harness is
connected at the
Main PCA
Reposition the
tubing
Replace the
damaged tubing
Replace Valve
Solenoid
Assembly
Replace the
Compressor
Replace the Valve
Body
Reconnect the
wiring harnesses
and retest
If problem
continues, install
old valve body
and new pilot
valves
Constant Audible
Alarm (no LED’s
on)
Fuse failure
Verify the continuity
of the fuse
Replace the fuse