EasyManua.ls Logo

Philips EverFlo Q - Page 22

Philips EverFlo Q
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 22 E
VERFLO
TM
SERVICE & TECHNICAL INFORMATION 1039055, VER. 06
EverFlo
Model
Normal Pressure for
units equipped with
Purge Valves
Normal Pressure for
units without Purge
Valves
Min
Max
Min
Max
1020000
1020001
1020002
1020003
1020002BR
1020003BR
1039362
1039363
NA
NA
15
27
1020004
1020005
NA
NA
NA
NA
1020006
1020009
1020010
1020011
1020012
1020016
1020017
1039366
1039368
1039370
16
24
14
25
1020007
1020008
1039367
12
22
12
23
1020013
NA
NA
11
22
1020014
1020015
1039364
1039365
NA
NA
14
24
9.
Record the peak pressure for four cycles.
CYCLE 1
CYCLE 2
CYCLE 3
CYCLE 4