Page 47 E
VERFLO
TM
SERVICE & TECHNICAL INFORMATION 1039055, VER. 06
Figure 8-5: TIE WRAP LOCATION
8.
Unclip the Main PCA shield from the Front Cabinet and push the PCA toward the bottom of the
cabinet until the power switch clears the front cabinet.
Figure 8-6: UNCLIP MAIN PCA
9.
Lift Main PCA out of the Front Cabinet.
Figure 8-7: MAIN PCA REMOVAL
10.
Remove the Power Cord wires from the J3 and J4 locations of the Main PCA.
11.
Use a large, flat bladed screwdriver to push in the tabs on the top of the strain relief and the bottom
strain relief to remove the Power Cord from the Rear Cabinet.