EasyManua.ls Logo

Philips EverFlo Q - Page 77

Philips EverFlo Q
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 77 E
VERFLO
TM
SERVICE & TECHNICAL INFORMATION 1039055, VER. 06
2. Slide the inlet boot bellows (see Section 8.2) onto the tee. Secure with the included cable tie. Cut any
excess cable tie and discard.
3. Insert the inlet tube assembly into the compressor housing making sure to orient the inlet boot bellows
on the same side of the compressor as the valve cover barb.
4. Secure the inlet tube assembly to the compressor housing by wrapping the two cable ties in the inlet
tube assembly repair kit around the elbow and tee and through the compressor housing cutouts. Cut any
excess cable tie using side cutters.

Table of Contents

Other manuals for Philips EverFlo Q

Related product manuals