38 Version: 1.2.0 MP125E N-470 Linear Actuator
10.1.3 Ambient Conditions and Classifications
Pay attention to the following ambient conditions and classifications for the N-470:
Area of application For indoor use only
Air pressure
Models not suitable for use in a vacuum:
1100 hPa to 0.1 hPa
Models N-470.xxV, .xxVY:
1100 hPa to 10
-6
hPa (high vacuum)
Models N-470.xxU, .xxUY:
1100 hPa to 10
-9
hPa (ultrahigh vacuum)
Highest relative humidity 80 % for temperatures up to 31 °C
Decreasing linearly to 50 % relative humidity at 40 °C
Operating temperature 10 °C to 40 °C
Storage temperature -20 °C to 75 °C
Transport temperature -20 °C to 75 °C
Maximum bakeout
temperature for N-470.xxV,
.xxVY, .xxU, .xxUY models
120 °C, for 12 hours, only in switched-off state
Overvoltage category II
Protection class I
Degree of pollution 1