EasyManua.ls Logo

POL-EKO RT 2014 - Alarms; Temperature; Humidity

POL-EKO RT 2014
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instruction manual RT 2014
POL-EKO-APARATURA sp.j. • ul. Kokoszycka 172 C • 44-300 Wodzisław Śl. • tel./fax +48/ 32 453 91 70 • e-mail info@pol-eko.com.pl
15
1) date,
2) time,
3) measurement from 1 sensor (T1),
4) battery voltage,
5) power supply,
6) name of file
9 ALARMS
Type of alarms depends on set parameters set in application AVIA
9.1 Temperature / humidity
User will be notified in case of:
Damage of sensor 1 or sensor 2 (indications transcend over the acceptable operating range of
sensor),
Values from sensor 1 or 2 transcend over the acceptable operating range set in parameters-
alarms event and the time set in column delay’ passed by.
Return of the value on sensor 1 or 2 to the set temperature and acceptable range.
If the delay’ is set e.g to 5 min (in case temperature will exceed), the alarm will not turn on if the
temperature value will return to acceptable range in faster than 5 minutes (see chart below).
By setting the delay properly you can prevent unnecessary alarms caused by e.g opening the
door to load the samples or defrosting function in the refrigerator.