EasyManua.ls Logo

PR 9113 - Configuration; Operating the Function Keys

Default Icon
65 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9113 - Product Version 9113-004 19
CONFIGURATION /
OPERATING THE FUNCTION KEYS
Documentation for routing diagram.
In general
When configuring the 9113, you will be guided through all parameters and you
can choose the settings which fit the application. For each menu there is a
scrolling help text which is automatically shown in line 3 on the display.
Configuration is carried out by use of the 3 function keys:
1 will increase the numerical value or choose the next parameter
2will decrease the numerical value or choose the previous parameter
3will accept the chosen value and proceed to the next menu
When configuration is completed, the display will return to the default state 1.0.
Pressing and holding
3 will return to the previous menu or return to the default
state (1.0) without saving the changed values or parameters.
If no key is activated for 1 minute, the display will return to the default state
(1.0) without saving the changed values or parameters.
Further explanations
Password protection: Programming access can be blocked by assigning a
password. The password is saved in the converter in order to ensure a high
degree of protection against unauthorised modifications to the configuration.
Default password 2008 allows access to all configuration menus.
Selection of units
By selection of temperature input you can choose which process units should
be shown in the display (see table). The process value is always displayed in
Celsius or Fahrenheit. This is selected in the menu point after selection of
temperature input.
CJC
IntheCJCmenuyoucanchoosebetweenCJCconnectorandinternalcoldjunction
compensation.TheCJCconnecter(PR5910Ex/PR5913Ex)mustbeordered
separately.

Table of Contents

Related product manuals