41
Advanced Configuration
5.1 Commands List
command
arguments answer on query comments
CTN
Continue (Measuring)
DRK
<n>(<x>)
n: Index of the lowe
st measuring
rate
x: lowest frequency in Hz, floating
point
“Dark reference”
take dark reference and save to flash
DNLD
<1..3>
-
Download spectrum. Currently for packet connections only.
ENC
<0..4> [<0..3>]
< -
2147483648
..
4294967295, ?>
encoder position
“Encoder Position“ :
$ENC <axis#> <function> <arg>
-index of axis
-optional: Function
-position (treated modulo 2^32)
Defined functions:
0: Set / Read Pos.
1: set count source <value> (0..9: A0, B0, A1, B1,
…; 10: SyncIn; 11..14: n.a., 15: Quardr.)
2: set preload value <value>
3: set preload event <value>
- Query currently supported for position only.
ETR
<func. Index>
<arguments>
see detailed description
“Encoder Trigger“,
see detailed description
IPCN
<0,1>, [eight
numbers
<0..255>]
-
Configure
TCP address and subnet mask. 1st arg: DHCP
on/off, args 2.5: fixed IP addr. args 6..9: mask (only if
DHCP off, i.e. 1st arg = 0)
LAI
<0…100, ?>
<value in %>
“Lamp intensity”
LED version: set on-time of LED between 1-
100% of the
exposure time.
NOP
<1..16>
<1..16>
Set number of peaks to evaluate.
WARNING: see detailed description below.
SCA
-
-
distance or thickness value [µm] for output value 32768