31
ingredient. It has been shown to be safe when used on the Premium V12
Series blood glucose meter.
To obtain disinfecting wipes and other information, please contact
Medlineat1-800-MEDLINE(1-800-633-5463)orvisitwww.medline.com.
Disinfecting Procedures
1. Take out one disinfecting wipe from the package and squeeze out any
excess liquid in order to prevent damage to the meter.
3. Remove the wipe. Allow the meter surface to dry completely.
4. Discard the used wipes.
Improper system cleaning and disinfection may result in meter
malfunction. If you have a question, please contact customer service at
1-866-469-2632 for assistance.
2. Wipe all exterior surface of the meter
including the display and buttons.
Hold the meter with the test strip slot
pointing down and wipe the area around
the test slot, be careful not to allow
excessliquidtogetinside.Keepthe
meter moist with disinfection solution
contained in the wipe for a minimum of
2 minutes. Follow the instructions on the
packagelabelofdisinfectingwipe.Use
two or more wipes if necessary.