32
This device has been validated to withstand 5,000 disinfection cycles
using the recommended disinfecting wipe/towelette. The tested number
of cycles is estimated by 1,000 tests per year on each meter and around
3 disinfection cycles per meter per day for 5 years, the expect life of
device.
Stop using the meter if you see any signs of deterioration such as the
LCDdisplayiscrackedorcloudy,thebuttonsnolongerfunction,orthe
outer casing is cracked. Please contact the customer service at 1-866-
469-2632 for a replacement meter if any of the signs of deterioration are
noticed.
The meter should be replaced after the validated number of disinfection
cycles or the warranty period, which ever comes rst.
NOTE:
•DoNOT clean or disinfect the meter while performing a test.
•Ifthemeterisbeingoperatedbyasecondperson,themeterandlancing
device should be decontaminated prior to handling by an assistant.
•DoNOT allow cleaning or disinfecting solution to get in the test slot, battery
compartment, or strip-ejection button.
•Ifyoudogetmoistureintheteststripslot,wipeitawaywiththecornerofa
tissue.
•Alwaysdrythemeterthoroughlybeforeusingit.
•Donotspraythemeterdirectlywithcleaningsolutionsespeciallythose
containingwater(i.e.soapywater),asthiscouldcausethesolutiontoenter
the case and damage the electronic components or circuitry.