OPTIONAL TEMPERATURE/HUMIDITY PROBES
2.5 mm stereo jack. Functional electrical isolation from PQube.
Optional probes plug into the PQube directly or through PSL-provided extension cable.
Typical: ±0.5ºC. Max: ±2ºC (-20 ~ +80ºC).
Typical: ±4.5% RH (20 ~ 80% RH), max: ±7.5% (0 ~ 100% RH).
Note: For optimal ambient temperature and humidity accuracy, use extension cable to avoid self-
heating of probe by PQube.
OPTIONAL CURRENT MEASURING MODULES
Pass-through (built-in current transformers)
20 amps RMS for CT4-20A, 50 amps RMS for CT4-50A
3.5 (±70 amps instantaneous for CT4-20A, ±350 amps instantaneous for CT4-50A)
12.8 kHz or 15.4 kHz (sampled at same rate as mains voltage measuring channels)
±0.2% rdg ±0.2% FS typical (10% ~ 120% FS). Every PQube factory tested at better than ±0.15% rdg
±0.15% FS.
0.34 inches (8.6 mm) max. diameter, 600 V UL-recognized insulation required
External current transformer
1 amp RMS for XCT4-1A, 5 amps RMS for XCT4-5A
3.5 (±3.5 amps instantaneous for XCT4-1A, ±17.5 amps instantaneous for XCT4-5A)
12.8 kHz or 15.4 kHz (measured at same rate as mains voltage measuring channels)
Accuracy - excluding external CT's
±0.2% rdg ±0.2% FS typical (10% ~ 120% FS). Every PQube factory tested at better than ±0.15% rdg
±0.15% FS.
Min. 20 AWG (0.52 mm2), max. 14AWG (2.1 mm2). 600 V UL-recognized insulation required
XCT5-0.333V-00, XCT5-1V-00, XCT5-5V-00, XCT5-10V-00, CTE1
External current transformer
0.333 V RMS, 1 V RMS, 5 V RMS, or 10 V RMS
3.5 (±1.17 Vpk, ±3.5 Vpk, ±17.5 Vpk, or ±35 Vpk)
12.8kHz or 15.4kHz (measured at same rate as mains voltage measuring channels)
Accuracy - excluding external CT's
±0.2% rdg ±0.2% FS typical (10% ~ 120% FS). Every PQube factory tested at better than ±0.1% rdg
±0.1% FS.
Min. 20AWG (0,52 mm2), Max. 14AWG (2,1mm2). 600V UL- recognized insulation required
External current transformer
0.333 V RMS, 1 V RMS, 5 V RMS, or 10 V RMS
OPTIONAL ATT1 VOLTAGE ATTENUATOR MODULES FOR ANALOG INPUT CHANNELS
±600 VDC/300 VAC to Earth
±1200 VDC/600 VAC to Earth
Nominal Measurement Range
Analog Input Channel Ratio
±0.2% reading typical at DC (>10% FS), plus error in analog input channels. For 50/60 Hz signals,
increase the analog input channel ratio by 0.1% (ATT1-0600V) or 0.4% (ATT1-1200V) per meter of cable
to approximately compensate for cable capacitance.