PQube User Manual 2.1
Page 69 of 113
OUTPUTS
Signal Relay
Connection
RLY1 PQube screw terminals [21], [29]. RLY2 PQube screw terminals [20], [28] with factory installed RLY
option. RLY3 PQube screw terminals [19], [27] with factory installed RLY option.
Rating
30 VAC/30 VDC, 300 mA max.
Function
Normally closed. Contacts open for duration of event or 3 seconds (whichever is longer).
Operate Time
20 milliseconds.
High Current Relay
Connection
RLY4 PQube screw terminals [17], [25]
Rating
30 VAC/30 VDC, 2A max.
Function
Normally closed. Contacts open for duration of event or 3 seconds (whichever is longer).
Operate Time
20 milliseconds.
COMMUNICATIONS
USB
Connection
Mini-B USB socket.
Future Applications
Future: USB mass storage device, and USB-based serial COM port.
Isolation
PQube provides at least 150VDC isolation to Earth (eliminates ground loops).
Optional Plug-in Ethernet Module
Connection
Standard RJ-45 socket (wired Ethernet).
Email
Sends emails after every event with data attached; user request real-time meters via e-mail, PQube
firmware upgrade via email, change PQube setup via email, incoming e-mail filters. Includes GIF graphs,
CSV spreadsheet files, PQDIF, HTML and XML summaries.
Web Server
Real-time meters. All events, trends and statistics recordings. Includes GIF graphs, CSV spreadsheet
files, PQDIF, HTML and XML summaries.
Modbus over TCP
Real-time meters with update rate of approximately 1 second. Event/trend-statistics counters can be
used for triggering downloads via FTP or web server.
FTP Server
File Transfer Protocol. Transfers files from PQube SD card to and from any computer. Limit: one
simultaneous connection.
SNTP
Simple Network Time Protocol for synchronizing PQube real-time clock to UTC. (2 second absolute -
UTC referenced).
CLOCK TIMING
Internal Real-Time Clock
Accuracy
Typical ±30 seconds/yr. Temperature compensated. ±120 seconds/yr. max drift
Optional SNTP (Requires ETH1)
Accuracy
±2 seconds absolute, UTC time.
OPERATING ENVIRONMENT
Ambient Conditions - Operating
-20°C ~ 50°C, 5% RH ~ 95% RH non-condensing
Transient Voltages
100kHz ring wave, 6 kV pk, IEC 61180, IEC 61000-4-5. Applied to voltage measuring terminals with
Performance Evaluation Class 1. (When applied to optional power supply mains terminal, supply’s fuse
may operate in PE Class 3 at test levels greater than 4 kV.)
EFT Burst Immunity
4 kV pk, IEC 61000-4-4, Performance Evaluation Class 1. Applied to power measuring terminals and
optional PS1 power supply mains terminals.
RF Field Strength Immunity
3V / m, IEC 61000-4-3 Test Level 2.
Magnetic Field Strength Immunity
30A / m, IEC 61000-4-8 Test Level 4.
Ingress Protection Rating (IP Rating)
IP20H, IEC 60529.