Trigger a Waveshape Change when the voltage
waveform changes abruptly. This is useful for
detecting power factor correction capacitor
switching.
If the voltage change from one cycle to the next
exceeds the selected threshold, for the selected
duration or longer, a Waveshape Change will be
triggered.
Probe_1_Overtemperature_Events=ON
Probe_1_Undertemperature_Events=OFF
Probe_1_Undertemperature_Threshold_in_Deg_C=0
Probe_1_Overtemperature_Threshold_in_Deg_C=60
Probe_1_Temperature_Event_Hysteresis_in_Deg_C=2
Probe_1_High_Humidity_Events=OFF
Probe_1_Low_Humidity_Events=OFF
Probe_1_Low_Humidity_Threshold_in_Percent_RH=0
Probe_1_High_Humidity_Threshold_in_Percent_RH=1
00
Probe_1_Humidity_Event_Hysteresis_in_Percent_RH=
2
Probe_2_Overtemperature_Events=OFF
Probe_2_Undertemperature_Events=OFF
Probe_2_Undertemperature_Threshold_in_Deg_C=0
Probe_2_Overtemperature_Threshold_in_Deg_C=50
Probe_2_Temperature_Event_Hysteresis_in_Deg_C=2
Show pop ups, recent event screens, and event depth
duration based on temperature/humidity thresholds
from either channel.