7 Calibraon
7.1 Bleach Calibraon (Three-Point with Zero)
The SP-210 is calibrated using a three-point calibraon, i.e., the zero-point is calibrated with deionized water
rst, and then the slope is calibrated using a concentraon of 1% and 10% sodium hypochlorite standard
soluon.
1. Rinse and ll the sample cell with deionized water.
*
NOTE
*
In case of emergency, you can use ”no sodium hypochlorite” clean sample
water, such as drink water, but please use deionized water to redo the zero calibraon
as soon as possible.
2. Power on the SP-210 by pressing
.
3. The SP-210 displays and measures the concentraon of bleach. If deionized water is used, the mea-
sured value will be very low; the value of sodium hypochlorite should be close to zero. A relavely
low non-zero reading (e.g., 0.01 etc.) is not a problem.
4. Press Cal/Sys ( ) to launch the Bleach CALIBRATION screen (Figure 3).
5. Press Zero (
) to set the zero point.
6. If the calibraon succeeds, a checkmark (
) and instrucons for the slope calibraon will appear
(Figure 4).
7. Rinse the sample cell three mes with the 1% sodium hypochlorite standard. Fill the sample cell with
the 1% standard.
8. Press Slope (
or ) to start the low-point slope calibraon.
9. If the calibraon succeeds, a checkmark (
) and a ”Calibraon Success” message will appear (Figure
5). Otherwise, a warning message is displayed.
10. Press Connue (
or ) to connue to the high-point slope calibraon (Figure 6).
11. Rinse the sample cell
three mes with the 10% sodium hypochlorite standard. Fill the sample cell
with the 10% standard.
12. Press Slope (
or ) to start the high-point slope calibraon.
13. If the calibraon succeeds, a checkmark (
) and a ”Calibraon Success” message will appear (Figure
7). Otherwise, a warning message is displayed.
14. Calibraon is now complete. Press Exit (
) to return to Bleach measurement mode.
*
NOTE
*
Aer calibraon is completed, the SP-210 will not automacally return to
Bleach measurement mode. If you press and hold Exit, the SP-210 will shut down
instead of returning to the measurement mode.
SP-210 User Manual service@pyxis-lab.com | +1 (866) 203-8397 7