EasyManua.ls Logo

Quorum Q150T E - Table 28 Metal Evaporation Profile Parameters

Default Icon
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Q150T Sample Preparation System
10473 - Issue 5 82 Q150T - Instruction Manual
Table 28 Metal Evaporation profile parameters
Name Default
Value
Minimum
Value
Maximum
Value
Comment
Material Carbon Material to be evaporated in this
process
Evaporation
Current (A)
0 0 50 Current applied during the
evaporation pulse
Evaporation time
(Minutes)
1 0.1 4 Duration of evaporation pulse
Out Gas Time
(Seconds)
10 2 60 Time source is outgassed
before evaporation pulse.
Out Gas Current
(A)
20 0 40 Current applied tosource at the
start of outgas step.
Thus allowing automated out
outgassing.
Note the current during the out
gas may be adjusted.
Max Out Gas
Current
(A)
40 0 50 Maximum current that can be
set during outgas
Max Out Gas
Voltage
(V)
10 0 20 The voltage across the rod is
limited to this value during
Outgas
Outgas source Yes No Yes Include or exclude outgassing
step
Tooling Factor 5 0.1 10 Tooling factor for this process
Pump hold No No Yes
If Yes, Q150T pumps down to
outgas-ready state and waits for
user input.
Note: If hold time (set in
System | Vacuum | Pump hold
time see page 46) is
exceeded, Q150T vents
chamber.
Flush chamber No No Yes
Set to Yes to flush chamber
with nitrogen before pumping
down.

Table of Contents

Related product manuals