Do you have a question about the R-Biopharm KOBRA CELL and is the answer not in the manual?
| Brand | R-Biopharm |
|---|---|
| Model | KOBRA CELL |
| Category | Measuring Instruments |
| Language | English |
Describes the primary applications and purpose of the KOBRA® CELL in HPLC analysis for aflatoxin detection.
Explains the electrochemical process involving bromine generation for aflatoxin derivatisation and its importance for detection.
Lists all the components included in the KOBRA® CELL package, such as the cell, power pack, and tubing.
Details the specific reagents (e.g., Distilled Water, Methanol, Potassium Bromide, Nitric Acid) needed for operation.
Information on optional or complementary products available for use with the KOBRA® CELL system.
Outlines the hazardous nature of mycotoxins and necessary precautions, including protective clothing.
Provides instructions for safely decontaminating labware and waste contaminated with hazardous substances.
Identifies and illustrates the individual components that make up the KOBRA® CELL assembly.
Key safety and handling instructions to ensure correct and leak-free assembly of the KOBRA® CELL.
Detailed instructions and visual guidance for the physical assembly of the KOBRA® CELL unit.
Notes on the KOBRA® CELL power pack, its design, and plug conversion for different regions.
Details the technical specifications of the power pack, including voltage, current, and battery life.
Diagrammatic representation of how the KOBRA® CELL connects within the HPLC system.
Step-by-step guide for connecting the KOBRA® CELL, mobile phase, and tubing within the HPLC setup.
Essential safety measures and checks to perform before initiating operation of the KOBRA® CELL.
How to turn the KOBRA® CELL power pack on and off for system operation.
Explanation of the LED indicator on the power pack and the conditions that trigger the error light.
Procedures for safely storing the KOBRA® CELL, including keeping the membrane wet.
Step-by-step instructions for daily cleaning to maintain the KOBRA® CELL and HPLC system.
Methods for regularly checking KOBRA® CELL performance to detect membrane deterioration.
Instructions on when and how to replace the KOBRA® CELL membrane for continued optimal performance.
Diagnoses potential causes when no peaks appear on the HPLC chromatogram, beyond the KOBRA® CELL.
Troubleshooting steps for low or absent peaks of Aflatoxin B1 and G1, checking connections and membrane.
Addresses issues related to abnormal peak shapes, such as poor resolution or tailing, often related to the column.
Explains why the error light on the power pack illuminates and checks to perform.
Addresses potential causes of leakage from the KOBRA® CELL, primarily related to screw tightness.
Troubleshooting steps for when the KOBRA® CELL causes high pressure readings on the HPLC system.
Guidance for issues with broken or lodged storage plugs, recommending professional repair.
Details the quality management systems (ISO 9001, ISO 13485) under which RBR products are developed.
Lists the support services offered to customers, including problem analysis and installation assistance.
Outlines the warranty terms and limitations for R-Biopharm Rhône Ltd products.