6.1 Storage
The KOBRA® CELL should be switched off and
disconnected for washing and storage (Fig. 5).
Always store the KOBRA® CELL filled with water in
order to keep the membrane wet.
Fig 4. Storage of KOBRA
®
CELL
To fill the KOBRA® CELL with water:
1. Fill a syringe with 5 - 10 ml of water and attach
the tip of the syringe to a pipette tip.
2. Unscrew one of the storage plugs / ferrules from
one inlet / outlet position of the KOBRA® CELL,
leaving the other inlet / outlet positions sealed
with storage plugs or ferrules.
3. Using the syringe with the pipette tip, insert this
into the empty inlet / outlet position on the
KOBRA® CELL and fill the cell with water until
water overflows from the syringe connection.
4. When finished, reinsert the storage plug / ferrule
to contain the water within the cell and keep the
membrane wet.
6
KOBRA
®
CELL Maintenance
Detector
HPLC Column
Out
Solvent waste
KC # 123
KC # 123
Connect
directly to the
detector
Dead end
ferrule to seal
port
KOBRA
®
CELL
waste
Inlet
Fig 5. KOBRA
®
CELL Switched Off for Washing and Storage