19
R-Biopharm Rhone Ltd
7.2 Reduced / No Aflatoxin B1
and G1 Peaks Observed
Check the KOBRA® CELL is connected correctly to
the HPLC (Fig 6).
Check the connections to the KOBRA® CELL from
the power pack.
Check the KOBRA® CELL membrane. The
membrane is the site of the actual derivatisation
reaction (ion-exchange occurs on the membrane).
The membrane starts a deep amber colour and,
through time and use, the centre of the membrane
becomes discoloured, eventually turning a pale
colour. The pale colour indicates that the
membrane will need replaced.
Check there is power in the power pack. Check
that the power pack has been turned on.
Check the internal components of the
KOBRA® CELL are in the correct order (Fig. 1), or
that there is not a blockage in the system.
No potassium bromide (or not enough potassium
bromide) was added to the mobile phase. Prepare
again.
No nitric acid was added to the mobile phase.
Prepare again.
Fig 9. Reduced / No Aflatoxin B1 and G1 Peaks Observed