LF
Multiplier
5.50
Equipment
required:
Item
Signa
[
Generator
T
piece
Terminating
pad
Coaxial
leads
Table
3 Item
No.
4
6
7
8
5.51
Connect
the
T
piece
to
the
'B'
Channel
input
of
the
UUT.
Connect
the
signal
generator
and
the
terminating
pad
to
the
T
piece.
5.52
Set
the
controls
of
the
UUT
as
for lows:
(1) N ORIVA L/BURST/CHECK
switch
to
NOR!vV\ L
(2) OPERA
TE/STANDBY
switch
to
OPERATE
(3)
LFselected
(4)
RESOLUTION
switch
to 1
Hz
5.53
Apply
the
frequencies
shown in
Table
11
at
a
level
of
8
mV
r.m.s.
(-29
dBm).
Verify
that
a
stable
and
accurate
display
is
obtained
at
this
level.
Increase
input
to
50
k
Hz
and
reduce
again
to
25
k
Hz.
Observe
mul tipl
ier
does
not
lock
to
intermed
iate
frequency.
TABLE
11
Frequency
Display
10
Hz
000010.00
Hz
25
kHz
025000.00
Hz
~
5.54
Remove
the
input
connection
to
the
UUT
and
check
that
the
display
does
not
exceed
5
counts.
5-13
9916
Vol.2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Racal Instruments 9916 and is the answer not in the manual?

Racal Instruments 9916 Specifications

General IconGeneral
BrandRacal Instruments
Model9916
CategoryMeasuring Instruments
LanguageEnglish