- 50 -
Use arrow buttons or to select the correct sample quantity.
For optional quantity variable (displayed value <0000>) enter any number using arrow buttons
Confirm selected sample quantity, confirmation message - <PLACExx PCS> - shall be seen.
Place the declared number of pieces in the container and when the result is stable (the symbol
is displayed) confirm the mass by pressing button,
The balance software automatically counts a single piece mass and enters the mode <PARTS
COUNTING> displaying the number of pieces which are on the pan (pcs). In the bottom line, a
single piece mass value is shown (if the option has been selected for <INFORMATION>
function).
CAUTION!
Remember that:
The total mass of all the pieces put on the weighing pan must not be greater than the balance maximum
weighing range.
The mass of a single piece cannot be lower than 0.1 reading unit of the balance. If this condition is not
met, the balance displays message: <Single part mass too low>.
While determining the number of pieces, wait for the stability pictogram , next confirm the quantity.
You can confirm the declared quantity by pressing button only after the stability pictogram is
displayed. Otherwise, the balance will not accept the measurement.
11.3.3. Setting the Reference Mass: Entering Mass Value
Procedure:
Press F button to which the function <ENTER SAMPLE> is assigned, wait to see the editing
window <ENTER SAMPLE>
Use arrow buttons to enter the known weight value of a single piece.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS PLUS Series and is the answer not in the manual?

RADWAG AS PLUS Series Specifications

General IconGeneral
DisplayLCD with backlight
Power Supply100-240 V AC 50/60 Hz
CalibrationInternal / External
InterfaceRS232, USB-A, USB-B
Operating Temperature+10° - +40°C
Pan Sizeø 85 mm
Repeatability0.1 mg
Linearity±0.2 mg
Stabilization time2-3 seconds