EasyManuals Logo

RADWAG AS PLUS Series User Manual

RADWAG AS PLUS Series
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #94 background imageLoading...
Page #94 background image
- 94 -
BEEP sound activated via BP command is inhibited if in-course of its activation the sound is activated by
means of other device: keypad, touch panel, proximity sensors.
Send all implemented commands
Format: PC CR LF
Command:
PC CR LF
- send all implemented commands
Response:
PC_A_”Z,T,S,SI…”
- command carried out, the terminal displays all
implemented commands.
Give balance type
Format: BN <CR><LF>
Response options:
BN_A_”x” <CR><LF>
- command understood, response: balance type
BN_I <CR><LF>
- command understood but not accessible at this moment
x series of types for a particular balance (in between inverted commas), with general balance type
in front
An example:
Command:
BN <CR><LF>
return balance type
Response:
BN_A_”AS”
balance type: ”AS R”
Give max capacity
Format: FS <CR><LF>
Response options:
FS_A_”x” <CR><LF>
- command understood, response: Max capacity
FS_I <CR><LF>
- command understood but not accessible at this moment
xMax value of reading units (in between inverted commas)
An example:
Command:
FS <CR><LF>
return Max capacity
Response:
FS_A_”220.0000”
Max capacity: ”220 g”
Give program version
Format: RV <CR><LF>
Response options:
RV_A_”x” <CR><LF>
- command understood, response: program version
RV_I <CR><LF>
- command understood but not accessible at this moment
xprogram version (in between inverted commas)
An example:
Command:
RV <CR><LF>
return program version
Response:
RV_A_” 1.1.1”
program version: ”1.1.1”
Set AUTOZERO function
Format: A_n <CR><LF>
Response options:
A_OK <CR><LF>
command carried out
A_E <CR><LF>
error in-course of command execution, no parameter or incorrect format
A_I <CR><LF>
command understood but not accessible at this moment
n parameter, decimal value determining autozero settings
n → 0 autozero off
1 autozero on
CAUTION!
Command changes settings for a current working mode.
An example:
Command:
A_1<CR><LF>
turn autozero function on
Response:
A_OK<CR><LF>
autozero function is on
AUTOZERO function operates until it is turned off by A 0 command.
Set ambient conditions state
Format: EV_n <CR><LF>
Response options:
EV_OK <CR><LF>
command carried out
EV_E <CR><LF>
error in-course of command execution, no parameter or incorrect

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS PLUS Series and is the answer not in the manual?

RADWAG AS PLUS Series Specifications

General IconGeneral
DisplayLCD with backlight
Power Supply100-240 V AC 50/60 Hz
CalibrationInternal / External
InterfaceRS232, USB-A, USB-B
Operating Temperature+10° - +40°C
Pan Sizeø 85 mm
Repeatability0.1 mg
Linearity±0.2 mg
Stabilization time2-3 seconds