- 93 -
An example:
Command:
OMG<CR><LF>
return current working mode
Response:
OMG_13_OK<CR><LF>
balance operates in Statistics working mode
Give accessible units
Command overview: Command returns units available for a particular device and for a current
working mode.
Format: UI <CR><LF>
Response options:
UI_”x
1
,x
2
, … x
n
”_OK<CR><LF>
command carried out, returns accessible units
UI_I <CR><LF>
command understood but not accessible at this
moment
x unit symbols, separated by means of comas
x → g, mg, ct, lb, oz, ozt, dwt, tlh, tls, tlt, tlc, mom, gr, ti, N, baht, tola, u1, u2
An example:
Command:
UI <CR><LF>
return available units
Response:
UI_”g, mg, ct”_OK<CR><LF>
response: available units
Set current unit
Command overview: Command sets current unit for a particular device.
Format: US_x <CR><LF>
Response options:
US_ x_OK <CR><LF
command carried out, response: current unit
US_E <CR><LF>
error in-course of command execution, no parameter or incorrect
format
US_I <CR><LF>
command understood but not accessible at this moment
x parameter, units symbols: g, mg, ct, lb, oz, ozt, dwt, tlh, tls, tlt, tlc, mom, gr, ti, N, baht, tola, msg,
u1, u2, next
CAUTION! if x=next, the command swaps to another avaialable unit on the list (it simulates „ ” button
pressing).
An example:
Command:
US_mg<CR><LF>
set „mg” unit
Response:
US_mg_OK<CR><LF>
„mg” set as a current unit
Give current unit
Command overview: command returns current unit.
Format: UG <CR><LF>
Response options:
UG_x_OK<CR><LF>
command carried out, response: current unit
UG_I <CR><LF>
command understood but not accessible at this moment
x parameter, unit symbol
An example:
Command:
UG<CR><LF>
return current unit
Response:
UG_ct_OK<CR><LF>
currently set unit is „ct”
Activate sound signal
Command overview: command activates BEEP sound signal for a specified amount of time
Format: BP_CZAS <CR><LF>
Response options:
BP_OK <CR><LF>
command carried out, BEEP sound signal activated
BP_E” <CR><LF>
no parameter or incorrect format
BP_I <CR><LF>
command understood but not accessible at this moment
TIME parameter, decimal value specifying how long shall the sound last, parameter given in [ms].
Recommended range <50 - 5000>.
If value greater than the permissible high limit is given, than BEEP sound is operated for the
maximum permissible amount of time.
An example:
Command:
BP_350<CR><LF>
activate BEEP for 350 ms
Response:
BP_OK<CR><LF>
BEEP activated
CAUTION!

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS PLUS Series and is the answer not in the manual?

RADWAG AS PLUS Series Specifications

General IconGeneral
DisplayLCD with backlight
Power Supply100-240 V AC 50/60 Hz
CalibrationInternal / External
InterfaceRS232, USB-A, USB-B
Operating Temperature+10° - +40°C
Pan Sizeø 85 mm
Repeatability0.1 mg
Linearity±0.2 mg
Stabilization time2-3 seconds