OMI <CR><LF>
n_”Nazwa modu” <CR><LF>:
n_”Nazwa modu” <CR><LF>
OK <CR><LF>
command carried out, response: accessible working modes
OMI_I <CR><LF>
command understood but not accessible at this moment
Mode name – parameter, working mode name, inserted in between inverted comas. The name takes
form given on a particular balance display, it is provided in a currently selected language.
n parameter, decimal value determining working mode number.
n →
1 Weighing
2
Parts Counting
3
Percent Weighing
4
Dosing
5
Formulas
6
Animal Weighing
8
Density of Solid Bodies
9
Density of Liquids
10
Peak Hold
11
Totalizing
12
Checkweighing
13 Statistics
CAUTION! Working modes numbering is identical for each kind of balance. The numbers are assigned to
working modes names. Some balances give only the number as a response.
An example 1:
Command:
OMI <CR><LF>
return accessible working modes
Response:
OMI <CR><LF>
2_”
Parts counting” <CR><LF>
4_”
Dosing” <CR><LF>
12_”Checkweighing” <CR><LF>
accessible working modes are given in
return:
mode number + name
OK <CR><LF>
command carried out
An example 2:
Command:
OMI <CR><LF>
return accessible working modes
Response:
OMI <CR><LF>
2 <CR><LF
>
4 <CR><LF>
12 <CR><LF>
accessible working modes are given in
return:
mode number
OK <CR><LF>
command carried out
Set working mode
Command overview: Command sets particular working mode.
Format: OMS_n <CR><LF>
Response options:
OMS_OK <CR><LF>
command carried out
OMS_E <CR><LF>
error in-course of command execution, no parameter or incorrect
format
OMS_I <CR><LF>
command understood but not accessible at this moment
n parameter, decimal value determining working mode number. To see detailed description go to
OMI command
An example:
Command:
OMS_13<CR><LF>
set Statistics mode
Response:
OMS_OK<CR><LF>
Statistics mode set
Give current working mode
Command overview: Command returns currently set working mode.
Format: OMG <CR><LF>
Response options:
OMG_n_OK <CR><LF>
command carried out, response: current working mode
OMG_I <CR><LF>
command understood but not accessible at this moment
n parameter, decimal value determining working mode number. To see detailed description go to
OMI command.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS PLUS Series and is the answer not in the manual?

RADWAG AS PLUS Series Specifications

General IconGeneral
DisplayLCD with backlight
Power Supply100-240 V AC 50/60 Hz
CalibrationInternal / External
InterfaceRS232, USB-A, USB-B
Operating Temperature+10° - +40°C
Pan Sizeø 85 mm
Repeatability0.1 mg
Linearity±0.2 mg