EasyManuals Logo

RADWAG AS R2 User Manual

RADWAG AS R2
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #89 background imageLoading...
Page #89 background image
- 89 -
Mode name parameter, working mode name, inserted in between inverted comas. The name takes
form given on a particular balance display, it is provided in a currently selected language.
n parameter, decimal value determining working mode number.
n →
1 Weighing
2
Parts Counting
3
Deviations
4
Dosing
5
Formulas
6
Animal Weighing
8
Solids Density
9
Liquids Density
10
Peak Hold
11
Totalizing
12
Checkweighing
13 Statistics
CAUTION! Working modes numbering is identical for each kind of balance. The numbers are assigned to
working modes names. Some balances give only the number as a response.
An example 1:
Command:
OMI CR LF
return accessible working modes
Response:
OMI CR LF
2_”
Parts counting” CR LF
4_”
Dosing” CR LF
12_”Checkweighing” CR LF
accessible working modes are given in
return:
mode number + name
OK CR LF
command carried out
An example 2:
Command:
OMI CR LF
return accessible working modes
Response:
OMI CR LF
2 CR
LF
4
CR LF
12 CR LF
accessible working modes are given in
return:
mode number
OK CR LF
command carried out
Set working mode
Command overview: Command sets particular working mode.
Format: OMS_n CR LF
Response options:
OMS_OK CR LF
command carried out
OMS_E CR LF
error in-course of command execution, no parameter or incorrect
format
OMS_I CR LF
command understood but not accessible at this moment
n parameter, decimal value determining working mode number. To see detailed description go to
OMI command
An example:
Command:
OMS_13 CR LF
set Statistics mode
Response:
OMS_OK CR LF
Statistics mode set
Give current working mode
Command overview: Command returns currently set working mode.
Format: OMG CR LF
Response options:
OMG_n_OK CR LF
command carried out, response: current working mode
OMG_I CR LF
command understood but not accessible at this moment
n parameter, decimal value determining working mode number. To see detailed description go to
OMI command.
An example:
Command:
OMG CR LF
return current working mode
Response:
OMG_13_OK CR LF
balance operates in Statistics working mode
Give accessible units
Command overview: Command returns units available for a particular device and for a current

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS R2 and is the answer not in the manual?

RADWAG AS R2 Specifications

General IconGeneral
Readability0.1 mg
DisplayLCD (backlit)
Tare Range-220 g
Linearity±0.2 mg
Protection ClassIP43
CalibrationInternal
InterfaceRS232, USB-A, USB-B