UI_”x
1
,x
2
, … x
n
”_OK CR LF
– command carried out, returns accessible units
– command understood but not accessible at this
moment
x – unit symbols, separated by means of comas
x → g, mg, ct, lb, oz, ozt, dwt, tlh, tls, tlt, tlc, mom, gr, ti, N, baht, tola, u1, u2
– response: available units
Command overview: Command sets current unit for a particular device.
– command carried out, response: current unit
– error in-course of command execution, no parameter or incorrect
format
– command understood but not accessible at this moment
x – parameter, units symbols: g, mg, ct, lb, oz, ozt, dwt, tlh, tls, tlt, tlc, mom, gr, ti, N, baht, tola, msg,
u1, u2, next
CAUTION! if x=next, the command swaps to another avaialable unit on the list (it simulates „ ” button
– „mg” set as a current unit
Command overview: command returns current unit.
– command carried out, response: current unit
– command understood but not accessible at this moment
x – parameter, unit symbol
– currently set unit is „ct”
Command overview: command activates BEEP sound signal for a specified amount of time
– command carried out, BEEP sound signal activated
– no parameter or incorrect format
– command understood but not accessible at this moment
TIME – parameter, decimal value specifying how long shall the sound last, parameter given in [ms].
Recommended range <50 ÷ 5000>.
If value greater than the permissible high limit is given, than BEEP sound is operated for the
maximum permissible amount of time.
– activate BEEP for 350 ms
BEEP sound activated via BP command is inhibited if in-course of its activation the sound is activated by
means of other device: keypad, touch panel, proximity sensors.