EasyManuals Logo

RADWAG AS R2 User Manual

RADWAG AS R2
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #90 background imageLoading...
Page #90 background image
- 90 -
working mode.
Format: UI CR LF
Response options:
UI_”x
1
,x
2
, … x
n
”_OK CR LF
command carried out, returns accessible units
UI_I CR LF
command understood but not accessible at this
moment
x unit symbols, separated by means of comas
x → g, mg, ct, lb, oz, ozt, dwt, tlh, tls, tlt, tlc, mom, gr, ti, N, baht, tola, u1, u2
An example:
Command:
UI CR LF
return available units
Response:
UI_”g, mg, ct”_OK CR LF
response: available units
Set current unit
Command overview: Command sets current unit for a particular device.
Format: US_x CR LF
Response options:
US_ x_OK CR LF
command carried out, response: current unit
US_E CR LF
error in-course of command execution, no parameter or incorrect
format
US_I CR LF
command understood but not accessible at this moment
x parameter, units symbols: g, mg, ct, lb, oz, ozt, dwt, tlh, tls, tlt, tlc, mom, gr, ti, N, baht, tola, msg,
u1, u2, next
CAUTION! if x=next, the command swaps to another avaialable unit on the list (it simulates „ ” button
pressing).
An example:
Command:
US_mg CR LF
set „mg” unit
Response:
US_mg_OK CR LF
„mg” set as a current unit
Give current unit
Command overview: command returns current unit.
Format: UG CR LF
Response options:
UG_x_OK CR LF
command carried out, response: current unit
UG_I CR LF
command understood but not accessible at this moment
x parameter, unit symbol
An example:
Command:
UG CR LF
return current unit
Response:
UG_ct_OK CR LF
currently set unit is „ct”
Activate sound signal
Command overview: command activates BEEP sound signal for a specified amount of time
Format: BP_CZAS CR LF
Response options:
BP_OK CR LF
command carried out, BEEP sound signal activated
BP_E” CR LF
no parameter or incorrect format
BP_I CR LF
command understood but not accessible at this moment
TIME parameter, decimal value specifying how long shall the sound last, parameter given in [ms].
Recommended range <50 ÷ 5000>.
If value greater than the permissible high limit is given, than BEEP sound is operated for the
maximum permissible amount of time.
An example:
Command:
BP_350 CR LF
activate BEEP for 350 ms
Response:
BP_OK CR LF
BEEP activated
CAUTION!
BEEP sound activated via BP command is inhibited if in-course of its activation the sound is activated by
means of other device: keypad, touch panel, proximity sensors.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG AS R2 and is the answer not in the manual?

RADWAG AS R2 Specifications

General IconGeneral
Readability0.1 mg
DisplayLCD (backlit)
Tare Range-220 g
Linearity±0.2 mg
Protection ClassIP43
CalibrationInternal
InterfaceRS232, USB-A, USB-B
Power Supply100 ÷ 240 V AC 50 / 60 Hz
Maximum Capacityup to 220 g

Related product manuals