04 Adjustment test
Mass comparison of internal adjustment weight with its value saved in balance memory
05 Internal weight correction
The function enables correcting the value of internal adjustment weight. The function is
disabled in verified balances
06 Automatic adjustment
Determination of factors triggering start of automatic internal adjustment:
0 no – none of factors will cause start of adjustment
1 time – adjustment triggered by time interval set in point 07
2 temperature – adjustment triggered by change of temperature
3 both – adjustment triggered by time and temperature
Function not available in verified balances.
07 Time of automatic adjustment
Determination of time interval, after which automatic adjustment process is initiated.
Function not available in verified balances.
Return to weighing mode
Changes introduced in balance memory will be saved on returning to
weighing with procedure of saving changes. Press ESC key for a few times
until the display shows a question: Save? As displayed, select one of
available options:
ENTER – save changes and go back to menu;
ESC – abandon changes and go back to menu.
(see point 7.1.2. Return to weighing)