8.4.3. Correction of Start Mass Expressed in Converter's Divisions
Enter factory setup submenu: <P.0.FAct / 0.3.CAL>.
Enter <0.3.4.Stu> submenu, value of start mass expressed in converter's
divisions is displayed.
Correct the value, to do it use operation panel, next press key to
confirm changes.
Exit to the home screen, to do it press key repeatedly.
8.5. Linearity Correction
Prior to linearity correction it is necessary to determine real scale
characteristics. Correction mechanism allows to enter corrections in 20 points
maximum.
8.5.1. Entering Linearity Correction Points
Enter factory setup submenu: <P.0.FAct / 0.4.Line / 0.4.1.dSG>, text
<Cont?> is displayed.
Press key to confirm. Text <Pnt1> is displayed (linearity connection
point no. 1).
Press key, window for entering mass value for linearity correction
point no. 1 is displayed.
Enter the given value, next press key to confirm, text <Pnt2>
is displayed (linearity connection point no. 2).
Press key, the program automatically suggests value for the next
linearity correction point.
Press key to confirm the suggested value, or enter a new value
(do it using the operation panel).
Keep repeating this procedure until you enter value close to maximum
capacity value.
While attempting to enter correction point value greater than
maximum capacity value, <Err Hi> error is signalled.