8.2. Factory Parameters List
No.
Name
Value
Description
P0.
FAct
-
Factory parameters
0.1.
Glob
-
Global parameters
0.1.1.
duu
-
Defining the weighing device.
0.1.2.
FAc
Serial number.
0.1.3.
tYP
1, 2, 4, 6, 7, 8, 9, 12
Weighing device type: 1 - WLC/A2;
2 - WLC/F, WLC/C2, 4 WTC,
6 Medical scale, 7 - Medical scale
(BMI disabled), 8 PUE C315;
9 PUE H315; 12 WLC/C/2.
0.1.4.
Gcor
0.9 - 1.1
Gravity correction factor.
0.1.7.
tSc
SLA, nInnH, no
Battery selection.
0.1.8.
CSt
nonE, d, A, V, b, SP, SC,
nt
Declaring the customer: nonE - none,
d - KERN, A - ADEMI, V - VWR,
b - BOECO, SP - Spectrum,
SC Schuller, nt NEW TECH.
0.1.9.
rtc
-
RTC synchronisation.
0.1.A.
ntE
YES, no
US market metrological requirements
activation.
0.2.
nnG
-
Metrology
0.2.1.
A/d
-
Converter's divisions.
0.2.2.
Uni
g, kg, lb
Adjustment unit.
0.2.3.
du1
0.0001, 0.0002, 0.0005,
0.001, 0.002, 0.005, 0.01,
0.02, 0.05, 0.1, 0.2, 0.5, 1,
2, 5, 10, 20, 50
Range 1 reading unit.
0.2.4.
dE1
no, 0.001, 0.01, 0.1, 1, 2,
5
Range 1 verification unit; no - non-
verified balance/scale.
0.2.5.
du2
0.0001, 0.0002, 0.0005,
0.001, 0.002, 0.005, 0.01,
0.02, 0.05, 0.1, 0.2, 0.5,
1, 2, 5, 10, 20, 50
Range 2 reading unit.
0.2.6.
dE2
no, 0.001, 0.01, 0.1, 1, 2,
5
Range 2 verification unit; no - non-
verified balance/scale.
0.2.7.
Ful
-
Maximum range + overload.
0.2.8.
rn2
-
Range switching point.
0.2.9.
uuE
-
External adjustment weight mass.
0.2.A
uui
-
Internal adjustment weight mass. „0”
- internal adjustment disabled.

Other manuals for RADWAG PUE H315