Entering correction values for linearity.
Additional modules activation
Wireless communication module: YES
- enabled, no - disabled.
IN/OUT module:YES - enabled, no -
disabled.
RS485 module: YES - enabled, no -
disabled.
Ethernet module: YES - enabled, no -
disabled.
4-20mA moduleYES - enabled, no -
disabled.
8.3. Device Defining
Defining the device at the production stage requires entering basic
balance/scale parameters: serial number, weighing device type, weighing
range.
Procedure:
Enter factory setup <P0.FAct>.
Go to <0.1.Glob / 0.1.1.duu> submenu, text <Cont?> is displayed.
Press key, text <nr FAc> is displayed, next you see window
for entering the serial number.
Enter the serial number, do it using the operation panel.
Press key to confirm, text <type> is displayed, next you see window
for balance/scale type selection.
Select balance/scale type, do it using operation panel (refer to factory
parameters table, section 8.2 of this manual).
Press key to confirm, weighing range window is displayed.
Select appropriate weighing range, do it using the operation panel.
Press key to confirm, <0.1.1duu> submenu is displayed.
Exit to home screen, to do it press key repeatedly.