Correction value ranges between 0.90000 ÷ 1.99999.
If the scale is adjusted in the place of use then the value
of gravitational coefficient, <0.1.4.Gcr> parameter, must be
1.00000. If the scale is adjusted away from the place of use
(longitudinal change) the value must be corrected.
9. OPTIONAL EXTENSION MODULES
9.1. Input/Output Module
The 637R module expands the PUE H315 indicator functionality by additional
4 inputs and 4 outputs. The module is intended to be mounted inside
the indicator. It is equipped with optoisolated inputs and semiconductor
outputs, and enables free configuration of both the inputs and the outputs
(using indicator menu). The indicator's housing cover features an additional
cable gland through which 3-metre cable 10 x 0.5 mm2 is fed, which cable
is terminated with numbered, isolated wires.
9.1.1. Technical Specifications
Output parameters
Output quantity
4
Output type
Solid-state relay
Cable cross-section
0.14 0.5mm
2
Maximum output current
0.5 A DC
Maximum output voltage
30 V DC, AC
Input parameters
Input quantity
4
Input type
Optoisolated
Cable cross-section
0.14 0.5mm
2
Input voltage range
5 24 V DC