MiniRAE 3000 User’s Guide
62
Datalog Type
The instrument has three datalog types:
Auto Default mode. Collects datalog information when the
instrument is sampling.
Manual Datalogging occurs only when the instrument’s
datalogging is manually started (see below for details).
Snapshot Datalogs only during single-event capture sampling.
Note: You can only choose one datalog type to be active at a time.
1. Press [N/-] to step from one option to the next.
2. Press [Y/+] to make your selection (the dark circle in the “radio
button” indicates “on”).
3. When you have completed your selection, press [MODE].
You will see two choices: Save and Undo. You have the opportunity to
register the new settings or to change your mind and revert to your
previous settings.
Press [Y/+] to save the changes.
Press [N/-] to undo the changes and revert to the previous settings.
Manual Datalog
When the instrument is set to Manual Datalog, you turn datalogging on
and off by stepping through the displays from the Main Display, and
then pressing the keys to select datalog on/off functions.
When you reach the screen that says “Start Datalog?” press
[Y/+] to start it. You see “Datalog Started,” confirming that
datalogging is now on.
When you reach the screen that says “Stop Datalog?” press [Y/+] to
stop it. You see “Datalog Stopped,” confirming that datalogging is now
off.