MiniRAE 3000 User’s Guide
7
WARNINGS
STATIC HAZARD: Clean only with damp cloth.
For safety reasons, this equipment must be operated and serviced
by qualified personnel only. Read and understand instruction
manual completely before operating or servicing.
Use only RAE Systems battery packs, part numbers 059-3051-000,
059-3052-000, and 059-3054-000. This instrument has not been tested
in an explosive gas/air atmosphere having an oxygen concentration
greater than 21%. Substitution of components may impair intrinsic
safety. Recharge batteries only in non-hazardous locations.
Do not mix old and new batteries or batteries from different
manufacturers.
The calibration of all newly purchased RAE Systems instruments
should be tested by exposing the sensor(s) to known concentration
calibration gas before the instrument is put into service.
For maximum safety, the accuracy of the instrument should be checked
by exposing it to a known concentration calibration gas before each
day’s use.
Do not use USB/PC communication in hazardous locations.