Determining The Lamp Type
The monitor can accommodate three lamp values: 10.6eV (standard),
9.8eV, and 11.7eV. The monitor automatically reads a marking on the
side of the lamp to set the proper Correction Factor. There are two
ways to determine the lamp type:
Remove the lamp and look for markings (bars) on the side:
• No bars: 10.6eV
• 1 bar: 11.7eV
• 2 bars: 9.8eV
Also, when the monitor is running, the lamp type is shown along with
the calibration and measurement gas and Correction Factor:
Note: This screen can be accessed from the reading screen by pressing
[N/-] four times.