Trying Out the Instrument
R&S
®
RTO2000
77Getting Started 1332.9719.02 ─ 10
6.13 Saving Data
After a measurement with the R&S RTO, you can save the resulting waveform
data for further evaluation or comparison. You can also save measurement
results, and device settings to repeat or restore previous measurements.
●
"Saving waveform data" on page 77
●
"Saving data of an acquisition series" on page 78
●
"Saving measurement results" on page 78
●
"Saving and restoring device settings" on page 79
Saving waveform data
1. Press the [Save Recall] key on the front panel (in the Setup area on the left).
2. Select "Waveform".
3. Check the "Source".
4. Set "Scope" to "Full Waveform".
Tip: If a cursor, zoom or measurement gate is defined, you can use these set-
tings to export only a part of the waveform.
5. Under "Save to file", tap "Save As".
6. The file selection dialog box shows the default storage directory:
C:\Users\Public\Public Documents\Rohde-Schwarz\RTx\
RefWaveforms
7. Tap the keyboard icon on the right of the "File Name" field.
8. Enter Waveform1 on the online keyboard.
9. Tap "ENTER" to close the online keyboard.
10. Select the file type: "*.bin".
11. Tap "Save".
The waveform data is saved to the files Waveform1.Wfm.bin and
Waveform1.bin in the selected directory.
Saving Data