EasyManua.ls Logo

Rangevision SPECTRUM - Page 24

Rangevision SPECTRUM
28 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
In case if not all marks are found on the image, you will also see the error message. It
may mean the following:
not all markers are visible on one of the snapshots,
not all markers are lit by the projector,
calibration plate is mounted at large angle to the scanner,
plate is too close or too far from the scanner - the image is not sharp,
the plate is damaged or dirty.
Please correct all deficiencies and press Capture.
Afterallthenecessarysnapshotsaretaken,waitfortheresultofcalibration procedure.
From now on the resulting accuracyofcalibrationwill bespecified on the home screen
inpixels.Agoodresult is accuracy no worse (not higher) than 0.1 pix. If you get a higher
result, repeat the calibration, carefully checking the plates position and brightness. Ifisis
necessary, repeat calibration withoutbluelightandfilters.
24

Other manuals for Rangevision SPECTRUM