MEASURE WITH MAM SENSOR
The MAM sensor is ready to measure dose and HVL for all anode/filter combinations without changing any settings.
Place the connected sensor centered in the field, with the crosshair 6 cm (4 cm in the US) from the examination
table front edge (A). The angle in the horizontal plane has neglectable impact on the measurement result.
For kVp measurements, select your anode/filter combination in the sensor settings, or by swiping right from the
home screen to access the quick settings.
NOTE! When measuring kV on Mo/Rh, use the included 2 mm Al filter. Follow the instructions
printed on the filter.
Expose.