SETTING: KVP MODE
The MAM sensor is ready to measure dose and HVL for all beam qualities without any settings. However, to get a
kVp reading you have to select a beam quality.
If your beam quality is not in the list, you have to change your anode/filter combination to one of the listed. Expose,
and you will get a kVp reading which is applicable to all anode/filter combinations for the mammography machine
(since the high voltage generator will behave identically regardless of selected anode/filter combination).
NOTE! This setting will only affect the kVp measurement.
RaySafe X2 User Manual
34