MO/RH KV WITH MAM SENSOR
Place the sensor centered and 6 cm from the chest wall. Place the 2 mm Al filter on the sensor. Be sure to cover
the active sensor area.
Recommended generator setting
≥ 100 mAs.
Select
Mo/Rh 2 mm Al in the sensor settings.
RaySafe X2 User Manual
30