MAS: MEASUREMENT PARAMETER DEFINITIONS
mAs is calculated from all recorded data.
mA is calculated as an average of all samples above 50% of peak. Rush currents are automatically removed. Live
readings are based on samples since last reading. For long measurements, the final reading is recorded
approximately 1–2 s before end trig.
Time starts the first time the current reaches 50% of the calculated mA and ends the last time it drops below 50%.
Live readings are time since start trig.
Pulses are counted every time the signal reaches trig.
Pulse rate and mAs per pulse are averages for the last 6 pulses.
PARAMETERS IN MA WAVEFORM
T: Time
P1, P2, P3: Pulses
mA: mA
: Displayed mAs value.
NOTE! If measurements are made simultaneously with a sensor and mAs cable, any shared
parameter (time, pulses or pulse rate) will be taken from the sensor.