Outputs
42 Rev. G Nov/2015 MI3
6.5 Thermocouple Output TC
Comm Box:
metal
Source:
object temperature
Signal:
TCJ, TCK, TCR, or TCS
Terminal:
TC, GND
This output can be configured as thermocouple output type J, K, R, or S. For that output, you must
install a dedicated compensation cable. The output impedance is 20 Ω. The output is short circuit
resistant.
The outputs <OUT1> and <TC> are not available at the same time!

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RayTek MI3 and is the answer not in the manual?

RayTek MI3 Specifications

General IconGeneral
Spectral Response8 to 14 µm
Response Time500 ms
Display Resolution0.1 °C
Accuracy±1% or 1°C (2°F), whichever is greater
EmissivityAdjustable 0.10 to 1.00