ton once. While paused, the “REC” symbol will stop ashing. Press
the Logger Button again to resume Datalogging, and the “REC”
symbol will begin to ash again.
3) Stop Datalogging; while the Datalogger function is paused, press
the REC Button for 2 seconds. The “REC” symbol will disappear
indicating the Datalogging function has stopped.
Manual Datalogger (Set sampling time = 0 seconds)
1) Set sampling time is to 0 second; Press the REC Button, and the
“REC” symbol will appear on the LCD screen. Press the Logger
Button, and the “REC” symbol will ash once and one Beep will
sound, at the same time the measuring data along the time infor-
mation will be saved in memory. The lower Display will show the
Position (record) number, which is saved as well.
2) To Change the Position Number; press the Down Button and the
lower position number will ash on the display. Press the Up or
Down Buttons to set the Position Number (1 to 99) to help identify
the measurement location. The lower Display will show P x (x = 1 to
99). After the position number is selected, press the Enter Button to
conrm.
3) Stop Datalogging; press the REC Button for 2 seconds, and the
“REC” symbol will disappear indicating the Datalogging function
has stopped.
Checking the Time
During normal functions (not Datalogging), press the Time Check Button
and the lower LCD display will show the Year, Month/Date, Hour/Minute.
Check Sampling Time Information
During normal functions (not Datalogging), press the Sampling Button and
the lower LCD display will show the Sampling Time Information.
SD Card Data Structure
1) The rst time the SD card is used in the meter, the SD card will
generate a folder: WBA01
www.reedinstruments.com